8
•
Troubleshooting
Vanquish Pumps (VH-P10-A-02) Operating Manual
Page 203
Message and Code
Possible Cause and Remedial Action
Code 4120 The rear seal wash
system has run out of wash
solution.
This message alerts you only when the drop counter functionality of the seal
wash detector is enabled (standard setting).
•
Verify that wash liquid is present in the seal wash reservoir.
•
Inspect the seal wash lines across the entire flow path for indications of
blockage or leakage. Verify that all seal wash lines are properly connected and
routed. Replace the seal wash lines as necessary (see
•
Inspect the peristaltic tubing for indications of blockage or damage. Replace
the tubing as necessary (see
•
Verify that the peristaltic tubing is properly inserted in the seal wash pump
and that the lever of the pump is not blocked.
•
Inspect the electrodes of the seal wash detector for contamination or damage.
Replace the seal wash detector as necessary (see
section 7.5.3, page 132
•
Start an additional seal wash cycle in Chromeleon (see
).
While the wash cycle is running, inspect the pump heads. If droplets leave the
pump head at the rear, the seal wash seal is defective. Replace the seal wash
seal (see
Code 4121 Piston seal leakage
has exceeded the recommended
limit.
This message alerts you only when the drop counter functionality of the seal
wash detector is enabled (standard setting).
Test the pump for piston seal leakage and take the recommend action (see
Code 4122 The rear seal leak
sensor detects drops constantly.
This message alerts you only when the drop counter functionality of the seal
wash detector is enabled (standard setting).
Inspect the electrodes of the seal wash detector for contamination or damage.
Replace the seal wash detector as necessary (see
Code 4125 Degasser malfunction The vacuum level monitoring function detected insufficient degasser vacuum.
Turn the pump off and on again. Check the degasser vacuum in Chromeleon.
After about 1 minute, the setting should change from
NotOk
to
Ok
. If the
degasser vacuum is still insufficient, the degasser may be leaky. Locate the
DegasserPressure
property in the
Command
window, write down the pressure
reading, which can provide helpful information to identify the source for the
problem, and contact Technical Support.
Code 4127 The pump drive is
still in undock position. Please
execute a dock command.
This message alerts you during pump maintenance (for example, when replacing
the piston seals or pistons) when you try to start the pump while the pistons are
not yet in the correct position for normal operation. To return the pistons into
the correct position, perform a
Dock
command.
Code 4148 Can't perform this
command while the flow is on.
You tried to run a self-test while the flow rate was not yet down to zero. Apply a
flow of 0 and repeat the command.
Содержание Vanquish VH-P10-A-02
Страница 1: ...Vanquish Pumps VH P10 A 02 Operating Manual 4820 4411 EN Revision 1 0 October 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...
Страница 238: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...