6
•
Operation
Vanquish Pumps (VH-P10-A-02) Operating Manual
Page 107
No.
Description
1
Mounting holes for mounting bracket
2
Mounting bracket
3
Mounting screws (Torx)
8. Insert the static mixer into the mounting bracket. Mind the direction
of flow through the mixer (indicated by the arrow on the static
mixer).
Figure 32: Static mixer
9. Connect the capillary mixer from the mixer kit to the port labeled
OUT
on the purge unit and to the static mixer inlet.
10. On the static mixer outlet, reconnect the capillary to the
autosampler injection valve.
Figure 33: Mixing system installed and connected
No.
Description
1
Capillary mixer on purge unit (
OUT
port)
2
Capillary mixer on static mixer inlet
3
Capillary to autosampler injection valve
No.
Description
1
Static mixer inlet
2
Static mixer outlet
Содержание Vanquish VH-P10-A-02
Страница 1: ...Vanquish Pumps VH P10 A 02 Operating Manual 4820 4411 EN Revision 1 0 October 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...
Страница 238: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...