c) Screw the filter top to the filter bottom.
Figure 69: Assembling the solvent line filter
5. Slide the filter holder onto the solvent line.
6. Retighten the reservoir cap hand-tight. Verify that the retaining
guide remains in the hole in the reservoir cap. If it does not, press
the retaining guide into the hole to ensure that the tubing is kept in
place in the cap.
7. To remove any air bubbles from the solvent lines, purge the pump.
No.
Description
1
Solvent line filter, bottom
2
Filter frit
3
Solvent line filter, top