6.9.2
Gradient Delay Volume, Dwell Volume, and Ripple
The gradient delay volume of an HPLC system is defined as the volume
of the entire flow path between the point where the gradient is formed
and the point where the mobile phase enters the column. Thus, the
gradient delay volume of an HPLC system is the volume that the pump
must deliver until a change in solvent composition reaches the column
inlet.
Figure 28: Gradient delay volume of an HPLC system
No.
Description
1
Pump with
1a
Point where the gradient is formed (purge unit)
2
Autosampler
3
Column
4
Detector
5
Gradient delay volume of the system
The pump contributes to the system gradient delay volume by the dwell
volume. For the dwell volume of the pump, see
section 9.1 Performance
Содержание Vanquish VH-P10-A-02
Страница 1: ...Vanquish Pumps VH P10 A 02 Operating Manual 4820 4411 EN Revision 1 0 October 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...
Страница 238: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...