6
•
Operation
Page 114
Vanquish Pumps (VH-P10-A-02) Operating Manual
Restarting the Pump
To restart the pump, follow these steps:
1. Engage the peristaltic tubing in the seal wash pump.
2. Reconnect the seal wash line and the solvent lines to the
appropriate reservoirs.
Retighten the reservoir caps hand-tight. Verify that the retaining
guide remains in the hole in the reservoir cap. If it does not, press
the retaining guide into the hole to ensure that the tubing is kept in
place in the cap.
3. Turn on the pump.
4. Prepare and restart the other modules in the Vanquish system,
following the instructions in the
Operating Manuals
for the modules.
Pay special attention to the
Preparing the Module for Operation
section.
5. Purge the pump and then have the pump deliver for a short time.
6. Let the system equilibrate and be sure that it is ready for operation
(see
section 6.6 Preparing the Pump for Operation, page 91
Содержание Vanquish VH-P10-A-02
Страница 1: ...Vanquish Pumps VH P10 A 02 Operating Manual 4820 4411 EN Revision 1 0 October 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...
Страница 238: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...