Maintenance and Service
Vanquish Pumps (VH-P10-A-02) Operating Manual
Page 135
2. Install the new Y-connector:
First, connect the Y-connector capillaries to the outlet check valves.
Then, connect the capillary to the purge unit.
3. Purge the pump using the solvent for your application.
4. Inspect the flow connections of the Y-connector for indications of
leakage. Tighten leaky connections when the system pressure is
down to zero.