7
•
Maintenance and Service
Page 142
Vanquish Pumps (VH-P10-A-02) Operating Manual
7.7.3
Replacing the Pistons
When
Damage of piston
Parts and additional items required
•
Pistons
•
Isopropanol
Preparations
Remove the pump head (see
Follow these steps
1. Remove the pistons from the pump head (or pump if necessary).
2. Observe the following:
♦
Write down the 3-digit piston calibration value (which is
imprinted on the rear of the piston, see picture) and the
installation position in the pump head (left or right). You might
need this assignment in a later step.
Figure 43: Piston calibration value
♦
Reinstallation of pistons
Be sure not to interchange the pistons.
♦
Installation of new pistons (replacement pistons)
New pistons can be installed in either position.
3.
When reinstalling a piston you removed
Clean the piston. Follow the steps in
4. To facilitate piston installation, inject a few drops of isopropanol into
the piston cavity in which you want to install the piston.
Содержание Vanquish VH-P10-A-02
Страница 1: ...Vanquish Pumps VH P10 A 02 Operating Manual 4820 4411 EN Revision 1 0 October 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...
Страница 238: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...