7
•
Maintenance and Service
Page 168
Vanquish Pumps (VH-P10-A-02) Operating Manual
7.9.2
Replacing the Inline Filter or Capillary Mixer
When
•
Impaired filter permeability
•
Damage of inline filter or capillary mixer, which connects the purge
unit to the inline filter
Parts and additional items required
•
As applicable
♦
Inline filter (static filter)
♦
Capillary mixer (volume: 25 µL), connecting the purge unit to
filter inlet
•
Solvent suitable for purging
Preparations
1. To flush out harmful substances, purge the pump with an
appropriate solvent.
2. To remove harmful substances from the components in the flow
path after the purge unit, have the pump deliver for a short time
with the solvent that was used for purging.
3. Turn off the pump flow and wait until the system pressure is down
to zero.
Follow these steps
1. Unplug the cable of the seal wash detector from the
DROP DET
port.
2. Unplug the cable of the right pump head from the
P-WORK
port if
applicable.
3. Remove the right Y-connector (see
4. As applicable, disconnect the capillary mixer on the filter inlet or
remove the capillary mixer on the filter inlet and on the purge unit.
5. On the filter outlet, disconnect the capillary that connects the inline
filter to the autosampler injection valve.
6. Pull the inline filter out of the mounting bracket.
Содержание Vanquish VH-P10-A-02
Страница 1: ...Vanquish Pumps VH P10 A 02 Operating Manual 4820 4411 EN Revision 1 0 October 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...
Страница 238: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...