5. Clean the internal wall surface of the acceleration assembly (Figure 5-
20).
Note
Note
Note
Note
Note. Ensure the acceleration nozzle is clean. If not, use a cotton
swab and cleaner to remove any contamination.
▲
6. Inspect the large diameter, impactor nozzle O-ring for damage or wear
(Figure 5-20). Replace it, if necessary. Apply a thin film of silicone
grease to the O-ring. Also, apply a light coating of silicone grease to the
aluminum threads of the upper acceleration assembly.
7. Using a general-purpose cleaner with a paper towel, clean the collector
assembly walls and plate (Figure 5-21).
Note.
Note.
Note.
Note.
Note. Most of the contamination in the inlet is usually found on
the collector plate.
▲
Collector plate
Vent tube
Weep hole
Figure 5-20.
Figure 5-20.
Figure 5-20.
Figure 5-20.
Figure 5-20.
Interior wall of inlet.
Figure 5-21.
Figure 5-21.
Figure 5-21.
Figure 5-21.
Figure 5-21.
Inside of inlet collector
assembly.