Figure 5-68.
Figure 5-68.
Figure 5-68.
Figure 5-68.
Figure 5-68.
Replace Sample Filter screen.
12. The Replace Sample Filter screen will display (Figure 5-68). Remove
the calibration filter with the calibration filter removal tool. Correctly
install and properly seat a new TEOM filter onto the mass transducer
using the regular filter exchange tool. (Follow the instructions for
changing a filter earlier in this section.) Select the
Next >
button.
Note.
Note.
Note.
Note.
Note. DO NOT use the calibration filter exchange tool for
installing or removing ANY filter other than the pre-weighed
calibration filter.
▲
Note.
Note.
Note.
Note.
Note. If the frequency doesn’t stabilize, the wizard will display a
screen with instructions to re-seat the filter. Re-seat the filter and
select the Next >
Next >
Next >
Next >
Next > button. The Stabilizing screen will display. If the
unit is unable find a stable frequency the second time, the unit
will fail the K0 audit (Figure 5-70).
▲