To program the instrument and begin collecting data:
To program the instrument and begin collecting data:
To program the instrument and begin collecting data:
To program the instrument and begin collecting data:
To program the instrument and begin collecting data:
1. Perform a leak check (refer to the following section for leak check
information).
2. Install a TEOM filter into the mass transducer (Section 5).
3. When in the System Status screen, ensure that the serial number listed
for the instrument matches the serial number on the back of the unit.
4. When in the Instrument Conditions screen, select the
Flows
button to
display the Flows screen. Select the
Flow Rates
button to select the
desired flow rates for the Main and Bypass flow channels. Select the
Flow Control
button to select the desired flow control (“Active” or
Passive”) and the desired standard and average temperatures and
pressures. (Refer to Section 4 for more information on screens and
settings.)
5. When in the Settings screen, select the
System
button, then the
Set
Time
button to set the current date and time. (Refer to Section 4 for
more information on screens and settings.)
6. When in the Settings screen, select the
Advanced
button, then the
Mass Transducer K0 Constant
button to confirm the current K0
setting of the TEOM unit. The number programmed into the unit must
match the K0 constant on the label near the mass transducer. (Refer to
Section 4 for more information on screens and settings.)
7. When in the Settings screen, select the
Data Storage
button to display
the Data Storage screen. Confirm the desired data is selected to be
logged by the instrument. (Refer to “Storing Data” later in this section
for more information on selecting storage variables.)
8. If you will be setting up the unit to receive an analog input, to transmit
analog outputs, or setting up the unit’s contact closure circuits, refer to
Section 4 for information on the screens and settings used for these
parameters.
9. If you will be using the password function to control access to the
unit’s operation, refer to Section 4 for information on setting up the
password function.