10. The Install Calibration Filter screen will display (Figure 5-67). Correctly
install and properly seat the pre-weighed calibration/verification filter
onto the mass transducer using the calibration filter exchange tool.
(Follow the instructions for changing a filter earlier in this section.)
Select the
Next >
button.
Note.
Note.
Note.
Note.
Note. DO NOT use the calibration filter exchange tool for
installing or removing ANY filter other than the pre-weighed
calibration filter.
▲
11. The Stabilizing screen will display while the instrument waits for the
frequency of the TEOM calibration filter. A countdown timer on the
screen will show the progress of the stabilization step (Figure 5-68).
When the stabilization “Complete” message displays, select the
Next >
button.
Note.
Note.
Note.
Note.
Note. If the frequency doesn’t stabilize, the wizard will display a
screen with instructions to re-seat the filter. Re-seat the filter and
select the Next >
Next >
Next >
Next >
Next > button. The Stabilizing screen will display. If the
unit is unable find a stable frequency the second time, the unit
will fail the K0 audit (Figure 5-70).
▲
Figure 5-67.
Figure 5-67.
Figure 5-67.
Figure 5-67.
Figure 5-67.
Install Calibration Filter screen.