Thermo Scientific recommends the following regular maintenance
procedures for the TEOM 1405:
Replace the TEOM filters
Replace the TEOM filter when the filter
loading percentage (displayed in the
Main screen) nears 100% or every 30
days.
Clean the sample inlet
Clean the sample inlet that is mounted
on the tripod each time that you replace
a TEOM filter (every 30 days).
Replace the in-line filters
Replace the mian flow in-line filter and
the bypass in-line filter every 6 months,
or as necessary.
Clean the air inlet system
Clean the air inlet system inside the mass
transducer once a year, or as necessary.
Rebuild the sample pump
Rebuild the sample pump once every 18
months, or as necessary. The pump
rebuild kit (59-008630) contains
instructions for rebuilding the pump.
These maintenance intervals are guidelines. Requirements for routine
maintenance are site-specific, and may vary from one location to another.