4. Mount the water trap filter assembly near the 1405 unit (Figure 2-4).
Figure 2-4.
Figure 2-4.
Figure 2-4.
Figure 2-4.
Figure 2-4.
Water trap filter assembly.
Water trap filter
Quick-connect fitting
Note.
Note.
Note.
Note.
Note. Drain the water trap as needed.
▲
5. From the 10m length of tubing for the bypass line, cut a piece of tubing
long enough to reach from the water trap filter to the bypass filter on
the back of the 1405 unit (Figure 2-3 and 2-4). Install the tubing into
the quick-connect fittings on the water trap filter and bypass filter.
6. Insert one end of the remaining section of bypass tubing into the quick-
connect fitting at the end of the coiled piece of tubing connected to the
water trap filter.
Note.
Note.
Note.
Note.
Note. The other end of the bypass tubing will be connected to the
bypass connection on the flow splitter when you assemble and
install the sample inlet (following section).
▲
7. Select a location on the pump tubing to install the vacuum gauge. It
should allow the gauge to be easily monitored (Thermo Scientific
suggests a location about half a meter from where the tubing is installed
in the pump connection on the back of the unit.) Cut the tubing and
install the two ends into the quick-connect fittings on the vacuum
gauge.
Note.
Note.
Note.
Note.
Note. Thermo Scientific strongly recommends that you use the
vacuum pump provided with the unit. If you choose to install a
different pump, it must be oil-free and able to maintain a 21" Hg
vacuum at a flow of 16.67 l/min.
▲