6. The Measure Flow screen will display.
Allow the flow to stabilize.
When
the flow is stable, press the flow button and enter the reading from the
flow device. Select the
Next >
button.
7. The Flow Audit Results screen will display the difference between the
flow rate of the instrument and the flow rate on the measurement
device (Figure 5-43). If the difference is less than 10 percent, you may
adjust the flow rate to reflect the value on the audit device. Select the
Yes
button to adjust the flow based on the results of the flow audit.
Select the
No
button to leave the original flow settings in place. Select
the
Next >
button to return to the Select Flow screen and audit
another flow channel.
8. The Select Flow to Audit screen will display again. The flow channel
that was just audited will be “grayed out” on the screen to show it was
audited during this session. If you want to audit another flow channel,
select another channel and follow steps 3 - 8 (and the wizard) to
complete additional flow channels. Otherwise, ensure that no flow
channel buttons are selected, and select the
Next >
button.
Note
Note
Note
Note
Note. If the difference is more than 10 percent, the flow will fail
the audit and the unit requires a leak check/flow calibration
(Figure 5-45).
▲
Figure 5-43.
Figure 5-43.
Figure 5-43.
Figure 5-43.
Figure 5-43.
Flow Audit Results screen.