T
HERMO
F
ISHER
S
CIENTIFIC
S
S
S
S
S
ECTION
ECTION
ECTION
ECTION
ECTION
5
5
5
5
5
M
AINTENANCE
AND
C
ALIBRATION
P
ROCEDURES
O
PERATING
G
UIDE
, TEOM 1405
5-33
3. The Select a Flow Audit Device screen will display. Select a flow audit
device. Select “Direct Flow Device” to audit the flow using a direct
flow measuring device (reading “l/min” adjusted for temperature and
pressure) such as the Streamline Pro. Select “FTS” to audit the flow
using the FTS system. FTS users will enter the device calibration
constant and the change in pressure from the FTS. Select the “Next >”
button.
4. The Select Flow to Audit screen will display. Select which flow to
audit. Press the “Next > “ button.
Note
Note
Note
Note
Note. If you selected the FTS option in the previous screen, you
will be prompted to enter the correct calibration constants before
proceeding with the audit (Figure 5-39).
▲
Figure 5-39.
Figure 5-39.
Figure 5-39.
Figure 5-39.
Figure 5-39.
Enter FTS Constants screen.