T
HERMO
F
ISHER
S
CIENTIFIC
S
S
S
S
S
ECTION
ECTION
ECTION
ECTION
ECTION
5
5
5
5
5
M
AINTENANCE
AND
C
ALIBRATION
P
ROCEDURES
O
PERATING
G
UIDE
, TEOM 1405
5-47
7. The Select Analog Output Range screen will display (Figure 5-57). Set
the voltage range for the analog output. Select the
Next >
button.
Note.
Note.
Note.
Note.
Note. Ensure that the voltage range selected for the analog output
channel matches the jumper voltage setting for the analog output
channel (step 3).
▲
8. The Connect Meter screen will display (Figure 5-58). Locate the test
point (Figure 5-54) for the analog output being calibrated and attach a
voltmeter set to VDC to the test point and to the ground test point on
the board. Select the
Next >
button.
Figure 5-57.
Figure 5-57.
Figure 5-57.
Figure 5-57.
Figure 5-57.
Select Analog Output
Range screen
Figure 5-58.
Figure 5-58.
Figure 5-58.
Figure 5-58.
Figure 5-58.
Connect Meter screen.