The flow splitter and sample tube are pre-assembled. Should it need to be
re assembled, follow the instructions below.
To set up the flow splitter assembly:
To set up the flow splitter assembly:
To set up the flow splitter assembly:
To set up the flow splitter assembly:
To set up the flow splitter assembly:
1. Locate the flow splitter (Figures 2-5 and 2-6).
2. Loosen the 1/2-inch sample tube fastener nut and slide the sample
tube down into the flow splitter so that the top of the installed sample
tube (or flow adapter) is approximately 15.5 cm (6") from the top of
the flow splitter (it must be between 5.75 and 6.25 inches from the top
of the flow splitter) (Figure 2-7).
3. Tighten the 1/2-inch sample tube nut. Ensure that the top of the flow
adapter remains between 5.75 and 6.25 inches from the top of the flow
splitter (Figure 2-7).
Figure 2-7.
Figure 2-7.
Figure 2-7.
Figure 2-7.
Figure 2-7.
Measuring the distance from the
top of the sample tube to the top
of the flow splitter.
Flow splitter
Bypass tube
Sample tube
1/2-inch sample tube nut
Figure 2-6.
Figure 2-6.
Figure 2-6.
Figure 2-6.
Figure 2-6.
Bottom of the flow splitter.