5
5
5
5
5
M
AINTENANCE
AND
C
ALIBRATION
P
ROCEDURES
O
PERATING
G
UIDE
, TEOM 1405
5-11
19. If the filter needs to be re-seated, open the door to the sampler and the
mass transducer and press straight down on the TEOM filter with the
bottom of the filter exchange tool (Figure 5-9). This ensures that the
filter is properly seated. Close the mass transducer and sensor unit
door. Select the
Next >
button
20. The system will again display the waiting screen while it is testing for a
stable frequency. If it still cannot obtain a frequency for the filter, it
will prompt the user to re-seat the filter a second time. If it still cannot
obtain a stable frequency, the procedure will prompt to replace the
filter or post a fail message (Figures 5-13and 14)
21. When the frequency stable, the system will display the Completing the
TEOM Filter Replacement Wizard screen (Figure 5-12). Select the
Finish >
button.
Figure 5-12.
Figure 5-12.
Figure 5-12.
Figure 5-12.
Figure 5-12.
Finish screen with successfully
completed message.