To confirm the system’s K
To confirm the system’s K
To confirm the system’s K
To confirm the system’s K
To confirm the system’s K
0
00
00
calibration:
calibration:
calibration:
calibration:
calibration:
1. Confirm that the K0 number entered into the instrument and the K0
number on the plate near the mass transducer are the same. The K0
number entered into the unit can be found in the Audit screen (Figure
5-35)
2. Ensure the instrument is at the normal operating temperature and
condition.
3. Ensure that the pre-weighed filter in the kit matches the humidity
conditions for the test, as shown on the card provided with the kit.
Note.
Note.
Note.
Note.
Note. If the filter does not match the conditions listed on the
humidity indicator, follow the instructions provided with the kit to
dry the filter to an acceptable level.
▲
4. In the TEOM Data screen, select the
Service
button to display the
Service screen, then select the
Calibration
button to display the
Calibration screen (Figure 5-33).
5. Select the
Mass Transducer K0 Verification
button to start the K0
Verification Wizard (Figure 5-62). Select the
Next >
button to begin
the procedure.
Figure 5-62.
Figure 5-62.
Figure 5-62.
Figure 5-62.
Figure 5-62.
Welcome to K0 Verification
Wizard screen.