To calibrate the main or bypass flow:
To calibrate the main or bypass flow:
To calibrate the main or bypass flow:
To calibrate the main or bypass flow:
To calibrate the main or bypass flow:
1. In the TEOM Data screen, select the
Service
button to display the
Service screen, then select the
Calibration
button to display the
Calibration screen (Figure 5-42).
2. Select the
Flow Calibration
button to start the Flow Calibration
Wizard (Figure 5-47). Select the
Next >
button to begin the procedure.
Figure 5-47.
Figure 5-47.
Figure 5-47.
Figure 5-47.
Figure 5-47.
Welcome to the Flow
Calibration Wizard screen.
3. The Select a Flow Audit Device screen will display. Select a flow audit
device. Select “Direct Flow Device” to audit the flow using a direct
flow measuring device (reading “l/min” adjusted for temperature and
pressure) such as the Streamline Pro. Select “FTS” to audit the flow
using the FTS system. FTS users will enter the device calibration
constant and the change in pressure from the FTS. Select the “Next >”
button.
4. The Select Flow to Calibrate screen will display. Select the
Calibrate
Main Flow
or
Calibrate Bypass Flow
button to calibrate the selected
flow. Select the
Next >
button.