When the 25-pin Phoenix Contact connector is attached to the unit with
the wired digital or analog device connections, the inputs/outputs must be
set up using the TEOM 1405 software screens:
When in the Instrument Conditions screen, select the
Analog Inputs
button to set up the analog inputs (1-4) and convert the incoming voltage
to the desired scale. The analog inputs are self-calibrating.
When in the Settings screen, select the
Analog & Digital Outputs
screen
to set up the analog outputs (1-8) and digital outputs (contact closures 1-
2):
•
When in the Analog & Digital Outputs screen, select the
Analog Outputs
button to set a minimum and maximum value for the output for the desired
output channel. Refer to Section 5 for information on calibrating the
analog outputs.
•
When in the Analog & Digital Outputs screen, select the
Contact Closure
button to select a variable, operator and compare value for the desired
contact closure channel (1-2).
Refer to Section 4 for complete information on setting the inputs/outputs
in the Analog Inputs and Analog & Digital Outputs screens.