Cleaning the Air Inlet
Cleaning the Air Inlet
Cleaning the Air Inlet
Cleaning the Air Inlet
Cleaning the Air Inlet
You must clean the heated air inlet in the TEOM 1405 once a year to
remove the buildup of particulate matter on its inner walls. You can order a
tapered bristle brush that is appropriate for cleaning the air inlet system
from Thermo Scientific. You will need a piece of plastic or another
protective material; soapy water, alcohol or freon solution; a 1/2-inch (or
adjustable) wrench and a soft brush to clean the air inlet.
Follow these steps to clean the air inlet system:
Follow these steps to clean the air inlet system:
Follow these steps to clean the air inlet system:
Follow these steps to clean the air inlet system:
Follow these steps to clean the air inlet system:
1. Turn off the TEOM 1405 unit.
2. Open the door of the unit (Figure5-30) and locate the thermistor in the
top of the mass transducer assembly.
3. Using the 1/2-inch wrench, remove the thermistor from the top of the
mass transducer assembly.
Note
Note
Note
Note
Note. The thermistor has a short thread depth. Installation/removal
should take 1 1/2 to 2 1/2 turns.
▲
4. Open the mass transducer (refer to the Replacing the TEOM Filter
section earlier in this chapter for instructions on opening the mass
transducer).
Figure 5-30.
Figure 5-30.
Figure 5-30.
Figure 5-30.
Figure 5-30.
Thermistor on top of mass
transducer.
Thermistors