, TEOM 1405
3-5
3. The Remove the TEOM filter screen will display. Remove the TEOM
filter from the transducer to ensure it isn’t damaged during the leak
check procedure. Select the
Next >
button.
4. The Disconnect Vacuum Line screen will display. Remove the main
vacuum line (pump) connected to the pump from the back of the unit
(Figure 3-3). Select the
Next >
button.
Pump connection
Figure 3-3.
Figure 3-3.
Figure 3-3.
Figure 3-3.
Figure 3-3.
Back of TEOM 1405 unit.
5. The Stabilizing screen will display. Allow 1 minute for the flows to
stabilize then select the
Next >
button.
6. The Reconnect Vacuum Line screen will display. Reinstall the pump/
vacuum tubing into the back of the unit. Select the
Next >
button.