A-4
Appendix
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Sample Conditions
Temperature
5 to 35 °C
Total Alkalinity
Less than 50 ppm CaCO
3
Sulfite
Sulfite cannot be present in the sample
Inlet Pressure
8 to 100 psig
Flow Rate
40 mL/minute nominal set by pressure regulator
Sample Inlet
1/4” NPTF tube fitting
Sample Drain
3/4” NPT male
Grab Sampler
Optional
Display Features
Type
Custom backlit LCD
Size
54 x 76 mm (2 1/8 in x 3 in)
Marquee
Temperature, operator prompts
Middle Line
Concentration, error codes
Lower Line
mV (selectable)
Text Display
Scrolling marquee for English language instructions
on setup, calibration and diagnostics
Calibration
DKA Calibration
Yes
DKA Calibration Points
3 points
Offline Calibration
Yes
Offline Calibration Points
1 point
Pre-programmed Standard Values
Yes
Custom Calibration
Programmable for concentration and volume
Operator Input
DKA, operator programmable, offline
Inputs (Potentiometric)
Ion Input
Chloride
Reference Input
Reference