Calibration
IV-9
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Step
Operator Action
Scrolling Marquee
Main Display
Notes
23
Press
to accept the new slope and
E
0
values
or
Press
to abort the calibration. The
new slope and E
0
values will not be saved.
The last saved calibration prior to the
aborted calibration will be used in the
measure mode.
NEW CALIBRATION SLOPE AND E
0
PRESS ENTER TO ACCEPT OR PRESS
EXIT TO ABORT WITHOUT SAVING
Calculated Slope
and E
0
If
is pressed,
refer to the
Calibration Abort
Steps
section.
24
Pull the diverter valve out.
Turn the inlet valve on.
Close the calibration port on
the flow cell cap.
PULL DIVERTER VALVE OUT
TURN INLET VALVE ON
CLOSE CAL PORT
PRESS ENTER
FLO On
25
Press
.
PULL DIVERTER VALVE OUT
TURN INLET VALVE ON
CLOSE CAL PORT
PRESS ENTER
FLO On
26
System will proceed to measure mode.
Actual temperature reading
Actual concentration
reading
System enters
measure mode, but
will remain on hold
until the hold timer
expires. Default hold
time is 30 minutes.