Calibration
IV-10
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Step
Operator Action
Scrolling Marquee
Main Display
Notes
Calibration Abort Exit Steps –
If
is pressed at any point during the calibration, proceed with the following steps
Exit 1
ARE YOU SURE?
PRESS ENTER TO ABORT OR PRESS
LAST SCREEN TO BACKUP
SUrE
?
Exit 2
Press
to abort the calibration.
or
Press
to return to the last screen
used in the calibration sequence prior to
when
was pressed. Continue to
follow the calibration steps according to
the marquee and main display directions.
ARE YOU SURE?
PRESS ENTER TO ABORT OR PRESS
LAST SCREEN TO BACKUP
SUrE
?
Exit 3
CALIBRATION ABORTED
PRESS ENTER
CAL
Abrt
Exit 4
Press
to abort the calibration.
CALIBRATION ABORTED
PRESS ENTER
CAL
Abrt
Exit 5
Pull the diverter valve out.
Turn the inlet valve on.
Close the calibration port on the
flow cell cap.
PULL DIVERTER VALVE OUT
TURN INLET VALVE ON
CLOSE CAL PORT
PRESS ENTER
FLO On
Exit 6
Press
.
PULL DIVERTER VALVE OUT
TURN INLET VALVE ON
CLOSE CAL PORT
PRESS ENTER
FLO On
Exit 7
No operator action needed.
The system will enter the measure mode
and remain on hold until the hold timer
expires (default hold time is 30 minutes).
Actual temperature reading
Actual concentration
reading
The system allows the operator to abort the calibration or return to the
calibration when
is pressed at any point during the calibration.
The following table lists the calibration abort steps with their marquee
display and required operator actions.
Calibration
Abort Steps