Calibration
IV-8
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Step
Operator Action
Scrolling Marquee
Main Display
Notes
17
Wait 0 to 8 minutes.
AWAITING DATA STABILITY
Actual readings
flashing
System continuously
evaluates the
data until a stable
condition is met.
If there is a calibration
error, the appropriate
error message will be
displayed.
18
Add standard 2 through the
calibration port.
Pipet 1.5 mL of the standard into the flow
cell by adding three increments of the
standard using the 0.5 mL pipet.
ADD STD2
PRESS ENTER
Add Std2
For calibration
standard 2, use the
29,600 ppm high level
chloride standard
(Cat. No. 211740HL).
19
Press
.
ADD STD2
PRESS ENTER
Add Std2
20
Wait 2 minutes.
MEASURING STD2
CIRCULATING
WAIT 2 MINUTES
Actual readings
flashing
System circulates
sample. Minutes will
be counted down on
the marquee.
21
Wait 0 to 8 minutes.
AWAITING DATA STABILITY
Actual readings
flashing
System continuously
evaluates the
data until a stable
condition is met.
If there is a calibration
error, the appropriate
error message will be
displayed.
22
No action necessary.
System calculates
new Eo and slope and
checks for errors.
If there is a calibration
error, the appropriate
error message will be
displayed.