Troubleshooting
VI-2
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
• Displays up to the last twelve calibrations using three screens:
1. Slope and E
0
(analyzers with a pH board or ISE board installed),
mV offset (analyzers with an ORP board installed) or
cell constant (analyzers with a conductivity board installed)
2. Date and Time
3. Temperature
• Press
to scroll between the three information screens for
each calibration
• Press
/
to view the last twelve calibrations
Calibration Log
Default
Operator Action
Scrolling Marquee
Main Display
Notes
Calibration Log
Press
while in the measure mode
to enter the diagnostics mode
or
or
CALIBRATION SLOPE AND E
0
# # . # (Slope)
# # . # (E
0
)
CALIBRATION OFFSET
# # # . # (mV offset)
CALIBRATION CELL CONSTANT
# . # # # (Cell Constant)
Press
to display the calibration
time and date
CALIBRATION TIME/DATE
HH:MM
MM/DD or DD/MM
Press
/
to scroll
through the last
12 calibrations
Press
to display the calibration
temperature and electrode response
CALIBRATION TEMP
# # . # c (Temperature)
Press
/
to scroll
through the last
12 calibrations
Press
to proceed to the next menu