Analyzer Operation
III-29
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Default
Operator Action
Scrolling Marquee
Main Display
Notes
TEST
TEST
(flashing)
SEL
SCrn
SEL CH1 or SEL CH2
on main display of
two channel analyzer
Press
to test the mA outputs (4-20)
and relays (rLY)
DIAGNOSTICS
appears above SETUP
in mode window
To Test tOUt:
4-20
Press
/
to scroll between
4-20 and rLY settings and select 4-20
SELECT mA OR RELAY
tESt
4 - 20
(flashing)
Press
to test 4-20 outputs
SELECT TEMP OR SENSOR
OUTPUT
4 - 20
tOUt or SOUt
(flashing)
Press
/
to scroll between
SOUt and tOUt settings and select tOUt
SELECT TEMP OR SENSOR
OUTPUT
4 - 20
tOUt
(flashing)
Press
to display the tOUt low value
Actual
low value
displayed
mA TEMP OUTPUT LOW VALUE LO
# . # c
4.0 mA are sourced at
output terminal 11
Press
to display the tOUt high value
Actual
high value
displayed
mA TEMP OUTPUT HIGH VALUE HI
# . # c
20 mA are sourced at
output terminal 11
Press
to accept the test and return
to the main setup mode
CH1 or CH2
(flashing)
SEL
SCrn
Display for two
channel analyzer only
Press
to return to the channel
specific menu options in the setup mode
Action for two
channel analyzer only
TEST
(flashing)
SEL
SCrn
SEL CH1 or SEL CH2
on main display of
two channel analyzer
Press
to test the mA outputs (4-20)
and relays (rLY)