Troubleshooting
VI-8
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
• The main display will show
0
and the marquee will display
KEYPAD
TEST PRESS ALL KEYS ONE AT A TIME
.
• Press all eight keys one at a time in any order.
* After the first key is pressed Error E033 will be displayed if a new key is
not pressed within 10 seconds.
Keypad Test
Default
Operator Action
Scrolling Marquee
Main Display
Notes
Keypad Test
KEYPAD TEST
PRESS ALL KEYS ONE AT A TIME
0
Press all keys one at a time in any order
KEYPAD TEST
PRESS ALL KEYS ONE AT A TIME
Display will show
numbers 1 through 8
as keys are pressed
After the first
key is pressed
error E033 will
be displayed
if a new key
is not pressed
within 10
seconds
No operator action needed
The system will enter the measure mode
if the keypad test is passed