Calibration
IV-11
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
Error
Operator Action
Scrolling Marquee
Main Display
Notes
Stability Time Out Error
Error
41
Press
to continue the calibration
despite the reading instability.
or
Press
to abort the calibration and
refer to Exit 3. Repeat the calibration or
refer to the troubleshooting section.
READINGS WERE UNSTABLE
PRESS ENTER TO CONTINUE
CALIBRATION OR PRESS EXIT
TO ABORT
E041
The required system
stability was not
reached and a time
out error occurred.
Perform electrode
cleaning, conditioning
and troubleshooting.
Calibration Errors
Error
42
Press
to abort the calibration and
refer to Exit 3.
Verify that the correct standards were
used during the calibration, the standard
values were entered correctly and in the
right order and the electrodes are working
properly and then repeat the calibration.
INVALID SLOPE
PRESS ENTER
E042
The calculated slope
was invalid.
Standard values were
likely entered in the
wrong sequence
or wrong standard
values were entered.
Error
107
Press
to abort the calibration and
refer to Exit 3.
CALIBRATION DATA TOO
CLOSE TOGETHER
PRESS ENTER
E107
Use new standards.
Perform electrode
cleaning, conditioning
and troubleshooting.
Error
109
Press
to continue the calibration
despite the bad slope.
or
Press
to abort the calibration and
refer to Exit 3.
CH1 BAD SLOPE
PRESS ENTER
E109
The calculated slope
was outside of the
recommended range.
Perform electrode
cleaning and
conditioning, refer to
the troubleshooting
section and repeat the
calibration.
At any point during a calibration, the appropriate error message will be
displayed if there is a calibration related error.
The following table lists common calibration error codes with their
marquee display and required operator actions.
Calibration
Error Codes