Preventive Maintenance and Servicing
Maintenance for the Sample Probe
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
5-9
Figure 5–2.
Filter Housing Lid Assembly
6.
Replace the large graphite gasket lid seal with a new one. Line up the
through hole in the lid seal with the hole in the lid.
7.
Inspect the 3 graphite gasket filter element seals and replace if
necessary.
8.
Install the retaining washer, the 3 graphite filter seals, and the new filter
element onto the assembly. See
Figure 5–2
for the correct order of
installation. Thread on the knurled filter lock screw and tighten until
snug.
9.
Before re-installing the filter housing lid assembly, the filter housing
itself can be cleaned using a clean cloth dampened only with clean
water.
10.
Re-install the filter housing lid assembly into the filter housing by
pushing it all the way in. Flip the retaining clamp back into position so
that it engages the small bar handle, and then turn the bar handle
clockwise to tighten. Turn the “T” handle approximately 1 turn
clockwise or until snug.
11.
Re-install the probe protection shield onto the probe. Close the two
pressure clamps to secure the shield to the probe assembly.
Содержание Omni FTIR Multi Gas CEMS
Страница 1: ...Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004 00 11Apr2013...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 10: ...About This Manual Where to Get Help vi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 20: ...Figures xvi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 22: ...Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 182: ......
Страница 206: ......
Страница 216: ......
Страница 234: ......