Preventive Maintenance and Servicing
Replacing the PC
5-56
Omni FTIR Multi Gas CEMS Instruction Manual
Thermo Fisher Scientific
1.
Shut the PC down through Windows, and then shut off the
PC/Console circuit breaker on the breaker box.
2.
Using the Phillips screwdriver, remove the blank panel in the front of
the system directly above the PC and console.
3.
Remove the screws holding the PC to the front rails. Slide the “Lock”
lever on the front of the PC console to the “Release” position. Slide the
PC and console out to gain access to PC cables.
4.
Disconnect the 9 cables from the rear of the PC (
Figure 5–46
). Note or
mark the location of each to facilitate re-connection. It is especially
necessary that the Ethernet, USB, and Serial (if used) connections are
made to the same ports they were disconnected from to avoid any issues
with driver software. A flat blade screwdriver may be necessary to
remove the large data cable connectors.
Figure 5–46.
PC Removal
5.
Slide the PC console back into the rack. Depress the tab at the rear of
each slide rail of the PC console to unlock it (
Figure 5–46
).
Содержание Omni FTIR Multi Gas CEMS
Страница 1: ...Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004 00 11Apr2013...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 10: ...About This Manual Where to Get Help vi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 20: ...Figures xvi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 22: ...Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 182: ......
Страница 206: ......
Страница 216: ......
Страница 234: ......