Calibration
Manual Background Spectrum Collection Procedure
4-8
Omni FTIR Multi Gas CEMS Instruction Manual
Thermo Fisher Scientific
Select tab Process and select Reprocess….
Verify or set the following. Resolution: 0.5, Apodization: Happ-Genzel,
Phase correction: Mertz, Zero filling: None, Final format: SingleBeam,
Correction: None, Saved spectral range: 4500 – 400. Press OK.