Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
D-1
Appendix D
Calogix Setup
This appendix provides a description of the Calogix software used to
control the temperature of heated sample lines and monitor the pressure in
the FTIR sample cell. For details, see the following topics:
●
“
Hardware Description
” on page
D-1
●
“
Finding the Correct COM Port
” on page
D-2
●
“
Setting the Virtual COM Port
” on page
D-4
●
“
Programming the Controllers and Setting Parameters
” on page
D-6
The pressure and temperature controller hardware is located in a 2U chassis
located in the rack. It consists of four 2-channel controllers and a base unit
(
Figure D–1
). The hardware is connected to the system computer with a
USB cable, with a type B connector on the controller unit and type A on
the computer. The USB cable should always be connected to the computer
at the lower-left USB socket (
Figure D–2
). The hardware is also connected
to a 220 VAC supply from circuit breaker panel.
Figure D–1.
Base Unit and Four Controllers
Hardware
Description
Содержание Omni FTIR Multi Gas CEMS
Страница 1: ...Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004 00 11Apr2013...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 10: ...About This Manual Where to Get Help vi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 20: ...Figures xvi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 22: ...Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 182: ......
Страница 206: ......
Страница 216: ......
Страница 234: ......