Preventive Maintenance and Servicing
Removing the Left Side Panel
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
5-33
7.
Re-install the purge gas generator by following the previous steps in
reverse order. Reset the regulator to its original pressure and lock it by
pushing down on the knob.
Note
Due to the effect of humidity on certain components inside the
FTIR, it is important to minimize the amount of time that the purge air is
off. There is a desiccant pack inside the FTIR, but if it will be off the purge
air for more than a few hours, an alternative source of purge air should be
connected.
▲
Use the following procedure to remove the side panel on the left side of the
system, in order to view or gain access to internal components.
1.
The side panel is held on the rack by posts and keyhole slots, therefore,
no tools are needed to remove it. Using the supplied key, unlock the
side panel. Simply grasp the hand grips on the panel and carefully lift it
about an inch, then pull the panel away from the rack to remove it
(
Figure 5–23
and
Figure 5–24
).
Figure 5–23.
Lock and Hand Grips
Removing the
Left Side Panel
Содержание Omni FTIR Multi Gas CEMS
Страница 1: ...Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004 00 11Apr2013...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 10: ...About This Manual Where to Get Help vi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 20: ...Figures xvi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 22: ...Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 182: ......
Страница 206: ......
Страница 216: ......
Страница 234: ......