Calibration
System Calibration Procedure Using IMACC Collect
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
4-11
Doubleclick on icon “ImaccCollect” and answer Yes when prompted
“Would you like to load “ftir_calibrate.collect?”.
Click on the configuration tab; click on config button of Spectrometer
Control.
Click the Collect tab to display the method and background used for the
calibration factors.
Verify or set the following.
Background: C:\backgrounds\bkg.spa;
Method: c:\methods\Omni method
Scan time: 0:56, Low Cutoff: 400, High Cutoff: 4500.
Use Method Shift Settings: Not enabled; Use Dynamic References:
Enabled;
Use Bias Correction: Not enabled; Use Dynamic Shifting: Enabled; Forget
Analysis State: Enabled
System Calibration
Procedure Using
IMACC Collect