Preventive Maintenance and Servicing
Replacing the National Instruments Boards
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
5-61
5.
Remove the card cage from its mount in the PC and rotate it up to
access the NI boards, being careful not to pull out any cables (
Figure 5–
51
).
Figure 5–51.
Positioning the Card Cage
6.
Using the Phillips screwdriver, remove the mounting screw of the NI
board to be replaced, and then gently remove the board from the edge
connector. (See
Figure 5–52
and
Figure 5–53
that show the Digital
Board.)
Figure 5–52.
Removing NI Board Mounting Screw