Troubleshooting
Service Locations
6-8
Omni FTIR Multi Gas CEMS Instruction Manual
Thermo Fisher Scientific
Check that the SYSTEM.STATUS value is showing the
expected mode; see
Table B–3
in Appendix B. Odd numbered
pins should have a continuous DC voltage supplied (Max 24
VDC).
●
Check wiring from terminal block to breakout board on
opposite side of electrical panel.
●
Check cable from breaker board to NI PCI-6709 board inside
PC.
●
Maintenance digital output is not reporting the expected mode
●
Check the key switch position. The vertical position is
maintenance mode and voltage should be conducted from the
odd terminal 25 to the even terminal 26.
For additional assistance, Thermo Fisher Scientific has service available
from exclusive distributors worldwide. Contact one of the phone numbers
below for product support and technical information or visit us on the web
at www. thermoscientific.com/aqi.