Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS
Establishing Communications
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
C-5
Table C–2.
Digital I/O Connections
Digital No.
Block Terminal
Description
1
1
Digital Input Ground
2
2
Gas Mode Input: Check Span
3
3
Gas Mode Input: Check Spike
4
4
Gas Mode Input: Check Zero
5
5
Gas Mode Input: Bkg & Flowback
6
6
Gas Mode Input: Blowback
7
7 & 8
Gas Mode Output: Check Span
8
9 & 10
Gas Mode Output: Check Spike
9
11 & 12
Gas Mode Output: Check Zero
10
13 & 14
Gas Mode Output: Background & Flowback
11
15 & 16
Gas Mode Output: Blowback
12
17 & 18
Gas Mode Output: Sample
13
19 & 20
FTIR Data Valid
14
21 & 22
Temperature Alarm
15
23 & 24
Maintenance Status Key Switch
16
25 & 26
N/C
17 27
N/C
18 28
N/C
Figure C–3.