IMACC Script Setting
Modbus Control Version
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
E-11
The script execution sequence is from left to right and top to bottom. The
main operation settings are included in the “while statement” loop. Initial
variable definitions are stated in the first two setting blocks on the left and
are named “Calc_2” and Calc_4”.
●
System Alarm Setting
This setting is included in the first container block inside the “while
statement” loop. The calc_8 setting is the definition of the system
temperature alarm.
If the gas cell temperature is less than 175 degC or the system temperature
alarm is triggered, the “alarm” would be flagged. Logically, there would be
three alarm output results based on the system temperature status. The
validation of CEMS measurement data is determined by the variable