Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
6-1
Chapter 6
Troubleshooting
This chapter provides the following troubleshooting and service support
information for the Omni FTIR Multi Gas CEMS:
●
“
Concentration Errors
” on page
6-3
●
“
Boot Up Error/IMACC Software Error
” on page
6-4
●
“
Temperature Errors
” on page
6-4
●
“
Cell Pressure Errors/Sampling System Errors
” on page
6-5
●
“
Communication Errors
” on page
6-6
●
“
Service Locations
” on page
6-8
The Technical Support Department at Thermo Fisher Scientific can also
be consulted in the event of problems. See “
Service Locations
” at the end of
this chapter for contact information. In any correspondence with the
factory, please note both the model number and serial number of the
instrument.
Use the following check sheet to verify that parameters and settings are
within specification. This will assist in diagnosing any problems with the
system.
Check Sheet
●
Visual Inspection of Operating Parameters:
●
Pressure on system regulator on right side of rack:
●
System pressure = 75 PSI
●
Pressures on pneumatic panel:
●
CO
2
Free Air = 2.0 – 3.0 bar (30 – 44 PSI)
●
Pump Air = 1.0 – 3.0 bar (12 – 44 PSI)
●
Flows on pneumatic panel:
●
FTIR Purge Air = 1.0-1.5 LPM
Содержание Omni FTIR Multi Gas CEMS
Страница 1: ...Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004 00 11Apr2013...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 10: ...About This Manual Where to Get Help vi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 20: ...Figures xvi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 22: ...Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 182: ......
Страница 206: ......
Страница 216: ......
Страница 234: ......