System Component Description
FTIR
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
7-9
display called a spectrum then appears on the computer’s screen. The
spectrum is a graphic representation of the sample in terms of light
absorbed (or transmitted) by the sample at various frequencies (wave
numbers) along the spectrum. The size and position of the peaks in the
spectrum are used to identify or quantify the composition of the sample.
Figure 7–6
shows the infrared beam path and data processing steps.
The major system components provide the basic system functionality for
the instrument.
The spectrometer compartment provides the electrical and optical
functions. The major components it includes are the interferometer,
source, and laser. Where possible, the optical components are pinned-in-
place for easy service and maintenance.
Figure 7–7.
Spectrometer Compartment
The FTIR analyzer uses Thermo Fishser Scientific’s EverGlo source. The
source unit can be replaced as needed. It is pre-aligned and needs no
adjustment when it is installed.
Optical Components
Source
Содержание Omni FTIR Multi Gas CEMS
Страница 1: ...Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004 00 11Apr2013...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 10: ...About This Manual Where to Get Help vi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 20: ...Figures xvi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 22: ...Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 182: ......
Страница 206: ......
Страница 216: ......
Страница 234: ......