Interfacing a DCS/PLC with the Omni FTIR Multi Gas CEMS
Establishing Communications
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
C-3
This section provides information on the methods available for retrieving
system data, accessing system status conditions, initiating system events,
and checking system calibration.
The 4-20mA analog output signals are labeled “ANALOG OUTPUT X1”
on the electrical panel assembly, which is accessible from the rear of the
rack. Two sets of terminals are labeled 1-11 for the individual outputs, with
the positive side on the right and the return side on the left. The positive
side terminal includes a disconnect feature so that externally connected
components may be disconnected from the current loop for local testing
and diagnostics.
The 4-20mA analog outputs are each capable of driving a loop resistance
down to 500 Ohms.
Refer to
Table C–1
for the analog output terminal assignments. Refer to
Modbus Protocol
Table B–1
, “Read Register” for Modbus methods for
retrieving data. Refer to
Figure C–2
to locate the pins on the rear panel
terminal connector.
Table C–1.
Analog Output Connections
Analog No.
Signal
1 SO2
2 NO
3 NO2
4 N2O
5 NH3
6 HCl
7 HF
8 CO
9 CO2
10 H2O
11 SF6
Establishing
Communications
Analog Outputs
Содержание Omni FTIR Multi Gas CEMS
Страница 1: ...Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004 00 11Apr2013...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 10: ...About This Manual Where to Get Help vi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 20: ...Figures xvi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 22: ...Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 182: ......
Страница 206: ......
Страница 216: ......
Страница 234: ......