Preventive Maintenance and Servicing
Replacing the Temperature Sensor
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
5-39
4.
Lift the power supply off of the base, disengaging it from the Velcro
underneath.
5.
Install the new power supply and re-install the cover panel by following
the previous steps in reverse order.
Use the following procedure to replace the FTIR RTD temperature sensor.
Equipment Required:
Wrench, 7/16-inch
Phillips screwdriver
Temperature sensor
CAUTION
High surface temperatures present. Allow sufficient time to
cool.
▲
1.
Shut down the FTIR heaters by turning off the Temperature Control
circuit breaker on the breaker box and allow it to cool for 1 hour. This
will also put the system in fail safe mode so that it will not be sampling
stack gas.
2.
Remove the left side panel per procedure “
Removing the Left Side
Panel
” for better accessibility (optional step).
3.
Using the Phillips screwdriver, remove the FTIR cover panel in the
front of the system.
4.
Slide the FTIR drawer out of the system to access the rear of the unit.
5.
At the rear of the FTIR cell, disconnect the temperature sensor
connector from the sensor by pulling it off, exposing the 3 metal
prongs (
Figure 5–30
).
Replacing the
Temperature
Sensor
Содержание Omni FTIR Multi Gas CEMS
Страница 1: ...Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004 00 11Apr2013...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 10: ...About This Manual Where to Get Help vi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 20: ...Figures xvi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 22: ...Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 182: ......
Страница 206: ......
Страница 216: ......
Страница 234: ......