IMACC Script Setting
CEMS_SampleMode
E-4
Omni FTIR Multi Gas CEMS Instruction Manual
Thermo Fisher Scientific
This setting is used to configure the data transmission commutation port.
They should be set by manufacturer before shipment to customers.
●
Spectrum collection setting
Following the Modbus multisensor setting is the script setting for the FTIR
optical bench, or interferometer. Double click the setting block named
“Collect EZ setup”, to set its setting subpanel to pop up, as shown.
On the top, there are six named function buttons. The setting panel is the
display under the “collect” button. A scan time of “0:58” is the duration of
an FTIR interferometer scan time for each data point. “Low Cutoff and
High Cutoff” define the low end and high end of the Mid IR spectrum
range. They are not recommended to be adjusted by users. The background
spectrum used for the absorption spectrum conversion is saved in the