Calibration
Manual Background Spectrum Collection Procedure
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
4-5
When running IMACC Monitor, press Stop Script and press OK go to tab
OMNIC.
In OMNIC, select Exp Set, tab Collect, and set the following. No. of
scans: 30, Resolution: 0.5, Final Format: Interferogram, Correction: None.
Manual Background
Spectrum Collection
Procedure