Calibration
System Calibration Procedure Using IMACC Collect
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
4-13
Click on the “...” button of Temperature Control and select
ModbusmultiSensor Control. Click on config button to display settings of
the Temperature Sensor.
Verify or set the following.
Optical Path: FrontMain, Type of data: IEEE float, Code: CellTemp,
Register:0x0a20,
COM Port: 6, Parity: None, Baud: 9600, Stopbits: 1, Databits: 8, Flow
Control: None
Press OK to close ModbusMultiSensor Settings box.
Click on the “...” button of Pressure Control and select
ModbusMultiSensor Control. Click on config button to display settings of
the Pressure Sensor.
Verify or set the following.
Optical Path: FrontMain, Type of data: IEEE float, Code: CellPressure,
Register:0x0a2a,
COM Port: 1, Parity: None, Baud: 9600, Stopbits: 1, Databits: 8, Flow
Control: None
Press OK to close ModbusMultiSensor Settings box.