Preventive Maintenance and Servicing
Temperature Control Box Components
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
5-65
4.
After replacing the internal components, install the cover. Place the
cover on the temperature control box with the lip closest to the front
panel. Line up the mounting holes and use the small Phillips
screwdriver to install the 8 mounting screws.
5.
Slide the temperature control box back in the rack. Using the Phillips
screwdriver, re-install the 4 front panel screws that secure the
temperature control box to the rack to complete the procedure.
6.
Turn on the Temperature Control circuit breaker on the breaker box to
complete the procedure.
Use the following procedure to replace any of the internal fuses in the
temperature control box.
Equipment Required:
Phillips screwdriver
Small Phillips screwdriver
Replacement Fuses
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal components.
▲
1.
Remove the temperature control box cover per the previously
mentioned procedure “
Removing and Installing the Cover
”.
2.
There are 16 fuses located on the PC Board inside the temperature
control box (plus 4 spare fuses.) To remove any one of them, simply
grasp the fuse with your fingers and pull it up and out of the connector.
Replacing the Internal
Fuses
Содержание Omni FTIR Multi Gas CEMS
Страница 1: ...Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004 00 11Apr2013...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 10: ...About This Manual Where to Get Help vi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 20: ...Figures xvi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 22: ...Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Страница 182: ......
Страница 206: ......
Страница 216: ......
Страница 234: ......